A calibration free ammonia breath sensor using a quantum cascade laser with WMS 2f/1f

K. Owen, A. Farooq
Applied Physics B, 116, pp. 371-383, (2014)

A calibration free ammonia breath sensor using a quantum cascade laser with WMS 2f/1f

Keywords

Quantum Cascade laser       

Abstract

The amount of ammonia in exhaled breath has been linked to a variety of adverse medical conditions, including chronic kidney disease (CKD). The development of accurate, reliable breath sensors has the potential to improve medical care. Wavelength modulation spectroscopy with second harmonic normalized by the first harmonic (WMS 2f/1f) is a sensitive technique used in the development of calibration-free sensors. An ammonia gas sensor is designed and developed that uses a quantum cascade laser operating near 1,103.44 cm−1 and a multi-pass cell with an effective path length of 76.45 m. The sensor has a 7 ppbv detection limit and 5 % total uncertainty for breath measurements. The sensor was successfully used to detect ammonia in exhaled breath and compare healthy patients to patients diagnosed with CKD.

Code

       DOI: 10.1007/s00340-013-5701-1

Sources

Website PDF

See all publications 2014